Enables concentrations of organic chlorinated solvents (e.g. tetrachloroethylene, trichloroethylene) present in wastewater to be easily measured.
*We also have the Waste water test kit WPT-132 / WPT-133 / WPT-135 which includes the Gas sampling pump GV-100 instead of the Plastic syringe pump.

Waste water test kit
Model | Substance to be Measured | Tube No. & Name | Measuring Range(mg/L) | No. of tests per box |
|
---|---|---|---|---|---|
WP-132 | Trichloroethylene | 132M | Trichloroethylene | 0.14 - 7.0 | 10 |
132LL | Trichloroethylene | 0.007 - 0.256 | 10 | ||
WP-133 | Tetrachloroethylene | 133M | Tetrachloroethylene | 0.13 - 6.3 | 10 |
133LL | Tetrachloroethylene | 0.005 - 0.174 | 10 | ||
WP-135 | 1,1,1-Trichloroethane | 135L | 1,1,1-Trichloroethane | 0.4 - 14 | 5 |
Substances to be measured and GASTEC standard detector tubes to be used
Substance to be Measured | Tube No. & Name | Measuring Range(mg/L) | No. of tests per box |
|
---|---|---|---|---|
Trichloroethylene | 132L | Trichloroethylene | 0.056 - 1.95 | 10 |
132LC | Trichloroethylene | 0.1 - 0.4 | 10 | |
Tetrachloroethylene | 133L | Tetrachloroethylene | 0.052 - 1.93 | 10 |
133LC | Tetrachloroethylene | 0.05 - 0.3 | 10 |
How to use
- 1.
- Measure 200mL of wastewater with the beaker, replace it in the glass jar, seal the opening with a stopper, and shake it for one minute.

- 2.
- Both ends of the detector tube are broken off, and it is inserted in the syringe or sampling pump (GV-100) for gas sampling.
- 3.
- The stopper of the glass jar is removed, and measurement is done with the point of the detector tube just above the surface of the liquid.

- 4.
- After the measuring time is over, the scale on the detector tube is read.
- 5.
- The temperature of the wastewater is measured.
- 6.
- By using the conversion data of the user manual, calculate the concentration in the wastewater.